Step 1
Draw 3 ml ACD-A for APC 20 in 20 ml syringe.
Draw 8 ml ACD-A for APC 60 in 60 ml syringe.
Step 2
Transfer 1 ml ACD-A from the 20 ml syringe into Plasma Chamber of Process Disposable (PD) for APC 20.
Transfer 2 ml ACD-A from the 60 ml syringe into Plasma Chamber of Process Disposable (PD) for APC 60..
Step 3
Draw venous blood using ACD-A syringe for APC 20 to the 3/4 oz mark (approx. 22 ml).
Draw 60 ml venous blood using ACD-A syringe for APC 60.
Step 4
Transfer total syringe volume into blood chamber of PD (red port).
Step 5
Load SmartPReP® with PD, close the lid, press the green start button.
NOTE: To prevent system imbalance, use correct balance weight when processing only one PD.
Step 6
After processing, use syringe with spacer and withdraw PPP (white port) until air bubbles are present.
Transfer contents to yellow cup on surgical field..
Step 7
Transfer contents to red cup on surgical field.60 ml syringe.
Transfer contents to red cup on surgical field.
Step 8
Mix APC+™ activator.
Transfer contents to clear cup on surgical field.